Application:

• Measurement of AMB in beverage grade carbon dioxide

Analyzer

The BevAlert Model 8900 provides direct measurement of Acetaldehyde, Methanol, and Benzene (AMB) in Carbon Dioxide.

The instrument is utilized by Specialty Gas Manufacturers and the Food and Beverage Industry to monitor trace impurities in CO2 used in carbonated beverages.

The BevAlert Model 8900 employs a photoionization detector (PID). The AMB in the gas sample are physically seperated using proprietary GC columns. A dual-column configuration with timed backflush to vent is used to strip off moisture and heavier gases. At sample injection, a fixed volume of sample is carried through the pre-cut column. The backflush is timed so that primarily the AMB and other similar compounds continue on to the analytical column. Contaminants are then backflushed to vent. Acetaldehyde, Methanol, and Benzene are separated from potentially interfering components on the analytical column and elute to the detector for analysis.

Features

• Direct measurement of Acetaldehyde, Methanol, and Benzene in Carbon Dioxide

• Interference free response

• Automatic baseline adjustment for long-term stability

• Automatic calibration

• Analog output ranges are user selectable

• Concentration and diagnostic alarm relays

• Graphic display of current or historical concentrations

• Multipoint sampling options

• RS-232 and optional LAN

 

Specifications

Analysis Time: Detector: PID (High-sensitivity PID Optional)
Column: Capillary
Oven Temperature: 100 °C, Nominal
Carrier Gas: Nitrogen, 15 cc/min, Nominal
Lower Detection Limit: Accuracy: 1% of Full-scale
Precision: 2% of Measured Value
Span Drift (24HR): <2% of Full-scale
Sample Flow Rate: 250-500cc/min, typical

Output:

Analog: (5) 0-20ma or 4-20ma loop power supplied, isolated.
Selectable for: gas concentration, unintegrated detector signal. Options for up to 16 additional programmable 0-20ma, 4-20ma or voltage outputs: 0-1V, 0-5V, or 0-10V.
Digital: RS-232, optional Local Area Network
Relays: (5) User programmable relays for concentration and diagnostic alarms (1A @ 30Vdc). Options for up to 32 additional relays available.
Inputs: Optional digital input board for 3 contact closure inputs. Supports start analysis, start calibration, and analyze calibration gas sample.