Application:

• Measurement of C1-C5 and Dimethyl Ether for oil & gas exploration

Analyzer

The PetroAlert Model 8900 combines the selectivity of gas chromatography with the sensitivity, broad dynamic range, and hydrocarbon selectivity of a flame ionization detector (FID) providing fast analysis of C1-C5 hydrocarbons and Dimethyl Ether required in Mud Logging for Oil & Gas exploration.

The PetroAlert 8900 C1-C5/DME analysis uses a dual injection technique; essentially, two separate chromatograms are combined, via dual injection valves, into a single chromatographic analysis. Each analytical train uses a timed backflushto-vent to strip off moisture and C6+ components. For the Dimethyl Ether analysis, the entire analytical column is set to backflush after the elution of Dimethyl Ether while the remainder of the C1-C5 analysis proceeds. The C1-C5 analysis uses a pre-cut column in series with the analytical column. Backflush of the pre-cut column is timed so that only the C1-C5 components are allowed to elute to the analytical column, while heavier components are diverted to a vent. C1-C5 hydrocarbons are then separated on the analytical column and elute to the flame ionization detector (FID) after the Dimethyl Ether analysis.

Features

• Analysis of C1-C5 Hydrocarbons and Dimethyl Ether for Gas Referencing
• Automatic baseline adjustment for long-term stability
• Automatic calibration to a known standard
• Analog output ranges are user selectable
• Multipoint sampling options
• RS-232 and optional LAN

Specifications

Analysis Time: <240 seconds
Detector: FID
Column: Packed
Oven Temperature: 50 °C, Nominal
Lower Detection Limit: <3ppm Dimethyl Ether, <10ppm C1-C5
Accuracy: ±5% (relative) within dynamic range.
Precision: 2% of Measured Value
Carrier: Hydrogen, 10cc/min typical
Fuel: Hydrogen, 30cc/min typical
Combustion Air: Air, 200cc/min typical

Output:

Analog: (1) 0-20ma or 4-20ma loop power supplied, isolated. Selectable for: gas concentration, unintegrated detector signal. Options for up to 20 additional programmable 0-20ma, 4-20ma or voltage outputs: 0-1V, 0-5V, or 0-10V.
Digital: RS-232, optional Local Area Network
Relays: (5) User programmable relays for concentration and diagnostic alarms (1A @ 30Vdc). Options for up to 32 additional relays available.
Inputs: Optional digital input board for 3 contact closure inputs. Supports start analysis, start calibration, and analyze calibration gas sample.